Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 20 de 87
Filtrar
1.
J Funct Biomater ; 15(7)2024 Jun 28.
Artigo em Inglês | MEDLINE | ID: mdl-39057299

RESUMO

The present study evaluated the mechanical behavior of five designs of Morse taper (MT) connections with and without the application of loads. For this, the detorque of the fixing screw and the traction force required to disconnect the abutment from the implant were assessed. A total of 100 sets of implants/abutments (IAs) with MT-type connections were used, comprising five groups (n = 20/group): (1) Group Imp 11.5: IA sets with a cone angulation of 11.5°; (2) Group SIN 11.5: with a cone angulation of 11.5°; (3) Group SIN 16: with a cone angulation of 16°; (4) Group Neo 16: with a cone angulation of 16°; and (5) Group Str 15: with a cone angulation of 15°. All sets received the torque recommended by the manufacturer. After applying the torque, the counter torque of the fixing screws was measured in ten IA sets of each group without the application of cyclic loads (frequencies ≤ 2 Hz, 360,000 cycles, and force at 150 Ncm). The other ten sets of each group were subjected to cyclic loads, after which the detorque was measured. Afterwards, the force for disconnection between the implant and the abutment was measured by traction on all the samples. The untwisting of the abutment fixation screws showed a decrease in relation to the initial torque applied in all groups. In the unloaded samples, it was found to be -25.7% in Group 1, -30.4% in Group 2, -36.8% in Group 3, -29.6% in Group 4, and -25.7% in Group 5. After the applied loads, it was found to be -44% in Group 1, -43.5% in Group 2, -48.5% in Group 3, -47.2% in Group 4, and -49.8% in Group 5. The values for the IA sets were zero for SIN 16 (Group 3) and Neo16 (Group 4), both without and with loads. In the other three groups, without loads, the disconnection value was 56.3 ± 2.21 N (Group 1), 30.7 ± 2.00 N (Group 2), and 26.0 ± 2.52 N (Group 5). After applying loads, the values were 63.5 ± 3.06 N for Group 1, 34.2 ± 2.45 N in Group 2, and 23.1 ± 1.29 N in Group 5. It was concluded that in terms of the mechanical behavior of the five designs of MT IA sets, with and without the application of loads, the Imp 11.5, SIN 11.5, and Srt 15 groups showed better results compared to the SIN 16 and Neo 16 groups, showing that lower values of cone angulation increase the friction between the parts (IA), thus avoiding the need to maintain the torque of the fixing screw to maintain the union of the sets.

2.
PLoS One ; 19(3): e0298462, 2024.
Artigo em Inglês | MEDLINE | ID: mdl-38457413

RESUMO

The present study compared three different implant and abutment sets of type Morse taper (MT) connection, with- and without-index, were analyzed regarding their mechanical behavior without and with cyclic load application simulating the masticatory function. Ninety implant and abutment (IA) sets were used in the present study, divided into three groups (n = 30 samples per group): Group A, Ideale solid straight abutment (one piece) without index; Group B, Ideale abutment with an angle of 30-degree (two pieces) without index; Group C, Ideale abutment with an angle of 30-degree (two pieces) with index. The abutment stability quotient (ASQ) values, detorque value and rotation angle were measured before and after the cycling load. Twenty IA sets of each group were submitted to mechanical load at 360,000 cycles. The ASQ without load were 64.7 ± 2.49 for the group A, 60.2 ± 2.64 for the group B, 54.4 ± 3.27 for the group C; With load were 66.1 ± 5.20 for the group A, 58.5 ± 6.14 for the group B, 58.9 ± 2.99 for the group C. Detorque values were lower in groups B and C compared to group A (p < 0.05). In conclusion, the presence of the index did not influence the stability values. However, solid straight abutments (group A) showed higher values of stability compared to groups of angled abutments (groups B and C).

3.
Heliyon ; 10(3): e25038, 2024 Feb 15.
Artigo em Inglês | MEDLINE | ID: mdl-38322837

RESUMO

Background: The goal of this in vitro study was to compare three different surfaces: two types of implant surfaces commercially available ([a] smooth/machined and [b] acid-treated surface) versus (c) anodized surface. Discs were manufactured with commercially pure titanium (CP) grade IV, which were subsequently analyzed by scanning microscopy and fibroblastic and osteoblastic cell cultures. Methods: Ninety-nine discs (5 × 2 mm) were manufactured in titanium grade IV and received different surface treatments: (i) Mach group: machined; (ii) AA group: double acid etch; and (iii) AN group: anodizing treatment. Three discs from each group were analyzed by Scanning Electron Microscopy (SEM) to obtain surface topography images and qualitatively analyzed by EDS. Balb/c 3T3 fibroblasts and pre-osteoblastic cells (MC3T3-E1 lineage) were used to investigate each group's biological response (n = 10/cellular type). The data were compared statistically using the ANOVA one-way test, considered as a statistically significant difference p < 0.05. Results: The AA group had numerous micropores with diameters between 5 and 10 µm, while nanopores between 1 and 5 nm were measured in the AN group. The EDX spectrum showed a high titanium concentration in all the analyzed samples. The contact angle and wetting tension were higher in the AA, whereas similar results were observed for the other groups. A lower result was observed for base width in the AA, which was higher in the other two groups. The AN showed the best values in the fibroblast cells, followed by Mach and AA; whereas, in the culture of the MC3T3 cells, the result was precisely the opposite (AA > Mach > AN). There was similar behavior for cell adhesion for the test groups (Mach and AN), with greater adhesion of Balb/c 3T3 fibroblasts compared to MC3T3 cells; in the AA group, there was greater adherence for MC3T3 cells compared to Balb/c 3T3 fibroblasts. Conclusions: The findings suggest that different surface characteristics can produce different biological responses, possibly cell-line dependent. These findings have important implications for the design of implantable medical devices, where the surface characteristics can significantly impact its biocompatibility.

4.
Medicina (Kaunas) ; 59(7)2023 Jun 22.
Artigo em Inglês | MEDLINE | ID: mdl-37512000

RESUMO

Background: The objective of this study was to evaluate the load transmitted to the peri-implant bone by seven different restorative materials in single-unit rehabilitations with morse taper implants using a strain gauge. Materials: In a polyurethane block that simulated type III bone, a morse taper platform implant was installed (3.5 × 11 mm) in the center and 1 mm below the test base surface, and four strain gauges were installed around the implant, simulating the mesial, distal, buccal and lingual positions. Seven similar hybrid abutment crowns were crafted to simulate a lower premolar using different materials: 1-PMMA; 2-glass ceramic over resin matrix; 3-PEEK + lithium disilicate; 4-metal-ceramic; 5-lithium disilicate; 6-zirconia + feldspathic; 7-monolithic zirconia. All groups underwent axial and oblique loads (45 degrees) of 150 N from a universal testing machine. Five measurements (n = 5) were performed with each material and for each load type; the microdeformation data underwent statistical analysis. The data were obtained in microdeformation (µÎµ), and the significance level was set at p ≤ 0.05. Results: There was no statistically significant difference in the evaluation among the materials under either the axial load or the oblique load at 45 degrees. In turn, in the comparison between axial load and oblique load, there was a difference in load for all materials. Conclusion: The restorative material did not influence the load transmitted to the bone. Furthermore, the load transmitted to the bone was greater when it occurred obliquely at 45° regardless of the material used. In conclusion, it appeared that the different elastic modulus of each material did not influence the load transmission to the peri-implant bone.


Assuntos
Coroas , Zircônio , Humanos
5.
Medicina (Kaunas) ; 59(7)2023 Jul 05.
Artigo em Inglês | MEDLINE | ID: mdl-37512062

RESUMO

Objective: The primary objective of the present retrospective clinical study was to evaluate and compare the clinical performance presented by castable abutments developed for the MT system versus intermediate machined abutments, specifically regarding prosthetic or implant fractures/loss; the secondary objective was to verify the looseness of the abutments and the behavior of the peri-implant soft tissues. Methods: This clinical retrospective study was conducted on patients rehabilitated between 2019 and 2020. Inclusion criteria were patients in good general health, with an implants-supporting single crown; with solid machined abutments (control group) or castable UCLA abutments; with a connection portion (base) machined in cobalt-chrome (test group) over Morse taper DuoCone implants in the posterior mandible area; and at least two years in function. Clinical assessment was carried out by the same professional, considering the following parameters: (A) prosthetic: (i) loosening of the fixation screw, (ii) fracture of the screw and (iii) the number of times the patient had some type of complication after the installation of the prostheses were evaluated; (B) biological: (i) without keratinized mucosa (KM), (ii) 1 mm or less, (iii) between 1 and 2 mm and (iv) greater than 2 mm of KM width; and the presence or absence of mucositis. Furthermore, radiographic evaluation was performed in order to assess the marginal bone loss. These evaluations permitted to compare the groups analyzed and patients enrolled. Data were statistically analyzed, with the level of significance set at α = 0.05. Results: 79 patients with 120 MT implants were evaluated (80 castable UCLA abutments and 40 machined solid abutments). The follow-up was from 2 to 4 years. There was a 100% implant survival rate. Therefore, the control group showed two fractured abutments (5%) and no abutment loosening (95% for prosthetic survival rate), whereas the test group showed no abutment fracture but nine loosening screws (11.3%) (100% for prosthetic survival rate). Keratinized mucosa was considered thin or absent in 19 implants in the control group (47.5%) and 42 in the test group (52.5%). Mucositis was found in 11 implants in the control group (27.5%) and 27 in the test group (33.8%). A positive correlation was observed between the width of keratinized mucosa and mucositis (r = 0.521, p = 0.002). The mean marginal bone loss was 2.3 mm, ranging from 1.1 to 5.8 mm. No correlation was observed when considering marginal bone loss versus the three parameters (implant diameter, implant length and time of the prosthesis in function). Conclusions: The results suggest that UCLA-type abutments are a viable option for rehabilitating implants with Morse taper connections, suggesting lower fracture risk. Further research is necessary to confirm these findings and thoroughly evaluate the clinical performance and long-term outcomes.


Assuntos
Mucosite , Humanos , Estudos Retrospectivos , Próteses e Implantes
6.
J Funct Biomater ; 14(3)2023 Feb 26.
Artigo em Inglês | MEDLINE | ID: mdl-36976052

RESUMO

The goal of this retrospective clinical study was to evaluate the behavior of Morse-taper indexed abutments by analyzing the marginal bone level (MBL) after at least 12 months of function. Patients rehabilitated with single ceramic crowns between May 2015 and December 2020 received single Morse-taper connection implants (DuoCone implant) with two-piece straight abutment baseT used for at least 12 months, presenting periapical radiograph immediately after crown installation were enrolled. The position of the rehabilitated tooth and arch (maxilla or mandible), crown installation period, implant dimensions, abutment transmucosal height, installation site (immediate implant placement or healed area), associated with bone regeneration, immediate provisionalization, and complications after installation of the final crown were analyzed. The initial and final MBL was evaluated by comparing the initial and final X-rays. The level of significance was α = 0.05. Seventy-five patients (49 women and 26 men) enrolled had a mean period of evaluation of 22.7 ± 6.2 months. Thirty-one implant-abutment (IA) sets had between 12-18 months, 34 between 19-24 months, and 44 between 25-33 months. Only one patient failed due to an abutment fracture after 25 months of function. Fifty-eight implants were placed in the maxilla (53.2%) and 51 in the mandible (46.8%). Seventy-four implants were installed in healed sites (67.9%), and 35 were in fresh socket sites (32.1%). Thirty-two out of these 35 implants placed in fresh sockets had the gap filled with bone graft particles. Twenty-six implants received immediate provisionalization. The average MBL was -0.67 ± 0.65 mm in mesial and -0.70 ± 0.63 mm in distal (p = 0.5072). The most important finding was the statistically significant difference comparing the values obtained for MBL between the abutments with different transmucosal height portions, which were better for abutments with heights greater than 2.5 mm. Regarding the abutments' diameter, 58 had 3.5 mm (53.2%) and 51 had 4.5 mm (46.8%). There was no statistical difference between them, with the following means and standard deviation, respectively, -0.57 ± 0.53 mm (mesial) and -0.66 ± 0.50 mm (distal), and -0.78 ± 0.75 mm (mesial) and -0.746 ± 0.76 mm (distal). Regarding the implant dimensions, 24 implants were 3.5 mm (22%), and 85 implants (78%) had 4.0 mm. In length, 51 implants had 9 mm (46.8%), 25 had 11 mm (22.9%), and 33 implants were 13 mm (30.3%). There was no statistical difference between the abutment diameters (p > 0.05). Within the limitations of this study, it was possible to conclude that better behavior and lesser marginal bone loss were observed when using abutment heights greater than 2.5 mm of transmucosal portion and when placed implants with 13 mm length. Furthermore, this type of abutment showed a little incidence of failures within the period analyzed in our study.

7.
Medicina (Kaunas) ; 59(1)2023 Jan 14.
Artigo em Inglês | MEDLINE | ID: mdl-36676792

RESUMO

Objectives: The present study compared two implants with different macrogeometries placed in healed alveolar sites, evaluating the insertion torque (ITV) and implant stability quotient (ISQ) values at three different periods. Methods: Seventy patients with a total of 100 dental implants were allocated into two groups (n = 50 per group): DuoCone implants (DC group) that included 28 implants in the maxilla and 22 in the mandible, and Maestro implants (MAE group) that included 26 in the maxilla and 24 in the mandible. The ITV was measured during the implant placement, and the ISQ values were measured immediately at implant placement (baseline) and after 30 and 45 days. Results: The mean and standard deviations of the ITV were statistically significant (p < 0.0001), 56.4 ± 6.41 Ncm for the DC group and 29.3 ± 9.65 Ncm for the MAE group. In the DC group, the ISQs ranged between 61.1 ± 3.78 and 69.8 ± 3.86, while the MAE group presented similar values compared with the other group, ranging between 61.9 ± 3.92 and 72.1 ± 2.37. Conclusions: The value of implant insertion torque did not influence the ISQ values measured immediately after implant placement. However, the ITV influenced the ISQ values measured in the two initial periods of osseointegration, with implants installed with lower torques presenting higher ISQ values.


Assuntos
Maxila , Osseointegração , Humanos , Torque , Maxila/cirurgia , Mandíbula , Coleta de Dados
8.
PLoS One ; 18(1): e0280684, 2023.
Artigo em Inglês | MEDLINE | ID: mdl-36662887

RESUMO

OBJECTIVE: The aim of the present in vitro study was to evaluate the resistance on quasi-static forces and in the fatigue mechanical cycling of a new implant design compared to two other conventional implant designs. MATERIALS AND METHODS: Eighty-eight implants with their respective abutments were tested and distributed into four groups (n = 22 per group): Morse taper connection implant (MT group), conventional external hexagon implant (EH con group), new Collo implant of external hexagon with the smooth portion out of the bone insertion (EH out group), and new Collo implant of external hexagon with the implant platform inserted to the bone level (EH bl group). All the sets were subjected to quasi-static loading in a universal testing machine, and we measured the maximum resistance force supported by each sample. Another 12 samples from each group were submitted to the cyclic fatigue test at 4 intensities of forces (n = 3 per force): 80%, 60%, 40%, and 20%. The number of cycles supported by each sample at each force intensity was evaluated. RESULTS: The three groups of implants with external hexagon connection had similar maximum strength values of the sets (p > 0.05). Meanwhile, samples from the MT group showed the highest resistance values in comparison to the other three groups (p < 0.05). In the fatigue test, the Collo out group supported a smaller number of cycles that led to the fracture than the other 3 groups proposed at loads of 80%, 60%, and 40%, and only at the load value of 20% all groups had the same performance. CONCLUSIONS: Within the limitations of the present in vitro study, the results showed that the new Collo implant performs better when installed at bone level.


Assuntos
Osso e Ossos , Implantes Dentários , Análise do Estresse Dentário , Dente Suporte
9.
J Appl Biomater Funct Mater ; 20: 22808000221119650, 2022.
Artigo em Inglês | MEDLINE | ID: mdl-35975914

RESUMO

A large number of materials with different compositions and shapes have been proposed and studied for the purpose of bone tissue regeneration. Collagen-based materials have shown promising results for this application, with improved physicochemical properties. The aim of the present in vivo animal study was to evaluate and compare two commercially available collagen-based biomaterials for bone regeneration, with these being implanted in circumferential bone defects created in the calvarium of rabbits. Twenty rabbits received bilateral parietal osteotomies, performed with the aid of a 6.5 mm diameter trephine. Two groups were created: the BC group, where the defect was filled with a scaffold composed of 90% bovine bone particles and 10% porcine collagen, and the EG group, where the defect was filled with a scaffold composed of 75% hydroxyapatite particles of bovine origin and 25% bovine collagen. Ten animals were sacrificed at 30 days and another 10 at 45 days after implantation, and the samples were processed and histologically analyzed. In the evaluations of the samples at 30 days, no important differences were found in the results. However, in the samples at 45 days after surgery, the EG group showed better results than the BC group samples, mainly in terms of the amount of bone matrix formation (P < 0.0001) and the volume in area measured in each sample, where the EG group had a value 65% higher than that in the BC group samples. Based on the results obtained, we conclude that the amount of collagen and the particle characteristics present in the composition of the scaffolds can directly influence the amount of neoformation and/or bone regeneration.


Assuntos
Materiais Biocompatíveis , Colágeno , Animais , Materiais Biocompatíveis/química , Regeneração Óssea , Bovinos , Colágeno/química , Durapatita , Coelhos , Crânio , Suínos
10.
Biomed Mater Eng ; 33(2): 101-112, 2022.
Artigo em Inglês | MEDLINE | ID: mdl-34511480

RESUMO

BACKGROUND: During the masticatory cycle, loads of different intensities and directions are received by the dental structures and/or implants, which can cause micromovements at the junction between the abutment and implant. OBJECTIVE: The objective of this in vitro study was to evaluate the behavior of three different implant connections subjected to different load values using a digital radiography system. Additionally, the torque values for removing the abutment screws were also measured and compared. METHODS: Ninety sets of implant and abutment (IA) were used, divided into three groups according to the type of connection (n = 30 per group): EH group, external hexagon type connection; IH group, internal hexagon connection; and, MT group, Morse taper connection. RESULTS: MT group showed the better vertical misfit behavior at the three intensity of load applied, in comparison with EH and IH groups. In the analysis of torque maintenance (detorque test), MT group showed higher values of detorque when compared with the measured values of EH and IH groups (p < 0.001). CONCLUSIONS: The IA sets of EH and IH groups showed a microgap in all levels of applied loads, unlike the MT group this event was not observed. In the detorque test, MT group increase in the torque values when compared to the initial torque applied, unlike EH and IH groups showed a decrease in the initially torque applied in all conditions tested. A positive correlation was detected between the misfit and detorque values.


Assuntos
Coroas , Implantes Dentários , Dente Suporte , Projeto do Implante Dentário-Pivô , Análise do Estresse Dentário , Próteses e Implantes , Torque
SELEÇÃO DE REFERÊNCIAS