Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 3 de 3
Filtrar
Mais filtros











Base de dados
Intervalo de ano de publicação
1.
Brachytherapy ; 19(4): 544-553, 2020.
Artigo em Inglês | MEDLINE | ID: mdl-32386884

RESUMO

PURPOSE: Dose escalation yields higher complete response to rectal tumors, which may enable the omission of surgery. Dose escalation using 50 kVp contact x-ray brachytherapy (CXB) allow the treatment of a selective volume, resulting in low toxicity and organs-at-risk preservation. However, the use of CXB devices is limited because of its high cost and lack of treatment planning tools. Hence, the MAASTRO applicator (for HDR 192Ir sources) was developed and characterized by measurements and Monte Carlo simulations to be a cost-effective alternative to CXB devices. METHODS AND MATERIALS: A cylindrical applicator with lateral shielding was designed to be used with a rectoscope using its tip as treatment surface. Both the applicator and the rectoscope have a slanted edge to potentially allow easier placement against tumors. The applicator design was achieved by Monte Carlo modeling and validated experimentally with film dosimetry, using the Papillon 50 (P50) device as reference. RESULTS: The applicator delivers CXB doses in less than 9 min using a 20375 U source for a treatment area of approximately 20 × 20 mm2 at 2 mm depth. Normalized at 2 mm, the dose falloff for depths of 0 mm, 5 mm, and 10 mm are 130%, 70%, and 43% for the P50 and 140%, 67%, and 38% for the MAASTRO applicator, respectively. CONCLUSIONS: The MAASTRO applicator was designed to use HDR 192Ir sources to deliver a dose distribution similar to those of CXB devices. The applicator may provide a cost-effective solution for endoluminal boosting with clinical treatment planning system integration.


Assuntos
Braquiterapia/instrumentação , Neoplasias Retais/radioterapia , Braquiterapia/métodos , Simulação por Computador , Desenho de Equipamento , Dosimetria Fotográfica , Humanos , Radioisótopos de Irídio/uso terapêutico , Método de Monte Carlo , Órgãos em Risco , Doses de Radiação , Dosagem Radioterapêutica , Planejamento da Radioterapia Assistida por Computador
2.
Brachytherapy ; 18(6): 852-862, 2019.
Artigo em Inglês | MEDLINE | ID: mdl-31327634

RESUMO

PURPOSE: The Bravos afterloader system was released by Varian Medical Systems in October of 2018 for high-dose-rate brachytherapy with 192Ir sources, containing new features such as the CamScale (a new device for daily quality assurance and system recalibration), channel length verification, and different settings for rigid and flexible applicators. This study mechanically evaluated the Bravos system precision and accuracy for clinically relevant scenarios, using dummy sources. METHODS AND MATERIALS: The system was evaluated after three sets of experiments: (1) The CamScale was used to verify inter- and intra-channel dwelling variability and system calibration; (2) A high-speed camera was used to verify the source simulation cable movement inside a transparent quality assurance device, where dwell positions, dwell times, transit times, speed profiles, and accelerations were measured; (3) The source movement inside clinical applicators was captured with an imaging panel while being exposed to an external kV source. Measured and planned dwell positions and times were compared. RESULTS: Maximum deviations between planned and measured dwell positions and times for the source cable were 0.4 mm for the CamScale measurements and 0.07 seconds for the high-speed camera measurements. Mean dwell position deviations inside clinical applicators were below 1.2 mm for all applicators except the ring that required an offset correction of 1 mm to achieve a mean deviation of 0.4 mm. CONCLUSIONS: Features of the Bravos afterloader system provide a robust and precise treatment delivery. All measurements were within manufacturer specifications.


Assuntos
Braquiterapia/instrumentação , Neoplasias/radioterapia , Calibragem , Desenho de Equipamento , Humanos , Radioisótopos de Irídio/uso terapêutico , Dosagem Radioterapêutica
3.
Brachytherapy ; 17(6): 1037-1044, 2018.
Artigo em Inglês | MEDLINE | ID: mdl-30122346

RESUMO

PURPOSE: Dose escalation to rectal tumors leads to higher complete response rates and may thereby enable omission of surgery. Important advantages of endoluminal boosting techniques include the possibility to apply a more selective/localized boost than using external beam radiotherapy. A novel brachytherapy (BT) rectal applicator with lateral shielding was designed to be used with a rectoscope for eye-guided positioning to deliver a dose distribution similar to the one of contact x-ray radiotherapy devices, using commonly available high-dose-rate 192Ir BT sources. METHODS AND MATERIALS: A cylindrical multichannel BT applicator with lateral shielding was designed by Monte Carlo modeling, validated experimentally with film dosimetry and compared with results found in the literature for the Papillon 50 (P50) contact x-ray radiotherapy device regarding rectoscope dimensions, radiation beam shape, dose fall-off, and treatment time. RESULTS: The multichannel applicator designed is able to deliver 30 Gy under 13 min with a 20350 U (5 Ci) source. The use of multiple channels and lateral shielding provide a uniform circular treatment surface with 22 mm in diameter. The resulting dose fall-off is slightly steeper (maximum difference of 5%) than the one generated by the P50 device with the 22 mm applicator. CONCLUSIONS: A novel multichannel rectal applicator for contact radiotherapy with high-dose-rate 192Ir sources that can be integrated with commercially available treatment planning systems was designed to produce a dose distribution similar to the one obtained by the P50 device.


Assuntos
Braquiterapia/instrumentação , Radioisótopos de Irídio/administração & dosagem , Planejamento da Radioterapia Assistida por Computador/instrumentação , Neoplasias Retais/radioterapia , Braquiterapia/métodos , Desenho de Equipamento/métodos , Dosimetria Fotográfica/métodos , Humanos , Método de Monte Carlo , Dosagem Radioterapêutica , Planejamento da Radioterapia Assistida por Computador/métodos , Reto/efeitos da radiação
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA