Effectiveness of epicardial atrial pacing using a bipolar steroid-eluting endocardial lead with active fixation in an experimental model
Bueno, Sávia Christina Pereira; Tamaki, Wagner Tetsuji; Silva, Marcelo Fiorelli Alexandrino da; Zambolim, Cristiane Maciel; Silva, Kátia Regina da; Martinelli Filho, Martino; Costa, Roberto.
Afiliação
  • Bueno, Sávia Christina Pereira; Universidade de São Paulo. Faculdade de Medicina. Hospital das Clínicas. Instituto do Coração. São Paulo. Brasil
  • Tamaki, Wagner Tetsuji; Universidade de São Paulo. Faculdade de Medicina. Hospital das Clínicas. Instituto do Coração. São Paulo. Brasil
  • Silva, Marcelo Fiorelli Alexandrino da; Universidade de São Paulo. Faculdade de Medicina. São Paulo. Brasil
  • Zambolim, Cristiane Maciel; Universidade de São Paulo. Hospital das Clínicas. Instituto do Coração. São Paulo. Brasil
  • Silva, Kátia Regina da; Universidade de São Paulo. Hospital das Clínicas. Instituto do Coração. São Paulo. Brasil
  • Martinelli Filho, Martino; Universidade de São Paulo. Hospital das Clínicas. Instituto do Coração. São Paulo. Brasil
  • Costa, Roberto; Universidade de São Paulo. Hospital das Clínicas. Instituto do Coração. São Paulo. Brasil
Acta cir. bras. ; 28(7): 543-550, July 2013. ilus, graf, tab
Article em En | VETINDEX | ID: vti-9009
Biblioteca responsável: BR1.1
Localização: BR68.1
ABSTRACT

PURPOSE:

To assess the effectiveness of bipolar epicardial atrial pacing using an active fixation bipolar endocardial lead implanted on the atrial surface in an experimental model.

METHODS:

A total of ten Large White adult pigs underwent pacemaker implantation under general anesthesia. Atrial pacing and sensing parameters were obtained at the procedure, immediate postoperative period and on the 7th and the 30th postoperative in unipolar and bipolar configurations.

RESULTS:

All procedures were successfully performed. There were no perioperative complications and no early deaths. Atrial pacing and sensing parameters for both unipolar and bipolar modes remained stable throughout the study. We observed a progressive increase in atrial thresholds, ranging from 0.49 ± 0.35 (at implantation) to 1.86 ± 1.31 volts (30th postoperative day), in unipolar mode. Atrial impedance measurements decreased slightly over time, ranging from 486.80 ± 126.35 Ohms (at implantation) to 385.0 ± 80.52 Ohms (30th postoperative day). Atrial sensing measures remained stable from the immediate postoperative period until the end of the study.

CONCLUSION:

The bipolar active fixation endocardial lead implanted epicardially can provide stable conditions of pacing and sensing parameters throughout the postoperative follow-up.(AU)